Shaoning Yao, Wei-Tsu Tseng, et al.
ADMETA 2011
We evaluated the potential and limitations of resonating nanomechanical microcantilevers for the detection of mass adsorption. As a test system we used mass addition of gold layers of varying thickness. Our main findings are: (1) A linear increase in mass sensitivity with the square of the mode number - a sensitivity increase of two orders of magnitude is obtained from mode 1 to mode 7 with a minimum sensitivity of 8.6 ag Hz-1 νm-2 and mass resolution of 0.43 pg at mode 7 for a 1 νm thick cantilever. (2) The quality factor increases with the mode number, thus helping to achieve a higher sensitivity. (3) The effective spring constant of the cantilever remains constant for deposition of gold layers up to at least 4% of the cantilever thickness. © IOP Publishing Ltd.
Shaoning Yao, Wei-Tsu Tseng, et al.
ADMETA 2011
A. Gupta, R. Gross, et al.
SPIE Advances in Semiconductors and Superconductors 1990
J.K. Gimzewski, T.A. Jung, et al.
Surface Science
Kenneth R. Carter, Robert D. Miller, et al.
Macromolecules